Ultraviolet Crosslinkers Operating Instructions and Service Manual IMPORTANT: Please read these instructions before operating your UVP UV
UV CROSSLINKERS UVP 10 14. Set the radiometer scale to 20mW/cm2. 15. Connect the radiometer to PEN 2 of the chart recorder. 16. Select 1V sc
UV CROSSLINKERS UVP 11 GENERAL SERVICING PROCEDURES A. Operations without Removing the Top Cover Changing the tubes It is recommended to ch
UV CROSSLINKERS UVP 12 Note: If more than one tube holder is to be replaced, the wiring schematic diagram may be of assistance. Replacement of st
UV CROSSLINKERS UVP 13 3. Release main connector by holding the nuts with a spanner and then on the outside back of the unit, release the two scr
UV CROSSLINKERS UVP 14 C. Operations with the Internal Chassis Assembly Removed Removal of the assembly 1. Release the wire to the micro swi
UV CROSSLINKERS UVP 15 Changing plastic feet The feet are fitted to the bottom of the cabinet. They are removed by unscrewing the screw mounted
UV CROSSLINKERS UVP 16 TECHNICAL SUPPORT UVP offers technical support on all of its products. If you have any questions about the product’s use,
UV CROSSLINKERS UVP 2 CONTENTS Introduction 3 Important Safety Information 3 Description and Specifications 3 Operation 4 A
UV CROSSLINKERS UVP 3 INTRODUCTION UVP Crosslinkers offer researchers an instrument to quickly, safely and efficiently expose samples to a contro
UV CROSSLINKERS UVP 4 • Internal safety interlock • Large LCD readout • Tactile membrane switch keypad • UV blocking viewing window • Large i
UV CROSSLINKERS UVP 5 The UV Crosslinker can be operated on the following settings: A. Preset UV Energy Exposure Setting: Push the PRESET a
UV CROSSLINKERS UVP 6 C. User-Defined UV Energy Exposure Setting: Sometimes it may be necessary for you to set your own standards for exposure
UV CROSSLINKERS UVP 7 APPLICATIONS The Crosslinker is a multi-purpose ultraviolet exposure instrument for use in the laboratory. A wide variety o
UV CROSSLINKERS UVP 8 CHANGING THE UV WAVELENGTH A UV Crosslinker is purchased with shortwave, longwave or midrange tubes. However, if user requ
UV CROSSLINKERS UVP 9 CROSSLINKER CALIBRATION PROCEDURE Equipment necessary: • Chart recorder • Two probes with banana plug adapters • One ra